Page 2 - 2019-James-Machine-OpManual-012419-final
P. 2

02

      Contents                                                  Introduction

                                                                Protect yourself from liability by testing products‘ static
                                                                coefficient of friction (COF) on your Michelman James Machine.
                                                                ASTM calls this test “the only method appropriate for testing
                                                                polishes for specific compliance with the D2047-93 standard.”
      INTRODUCTION.................................................02
                                                                To reproduce this real-life force, a piece of flooring material is
      SAFETY.................................................................03  placed on the table of the Michelman-James Machine and

      DATA CARD & CONTROLS...............................04     pressure is applied, and the point at which the testing foot of
                                                                the Michelman-James Machine begins to slip is the COF. This
      TEST MATERIAL SUPPLY LIST...........................05    value is displayed on the digital readout and recorded in
                                                                short-term memory.
      TEST MATERIAL PREPARATION......................06
                                                                Before Use
      MACHINE SETUP & PREP..................................07
                                                                The Michelman-James Machine is calibrated at the factory
      OPERATION........................................................08  (see page 10). Under normal use, no further calibration should
                                                                be necessary, unless a series of tests performed on a known
      TROUBLESHOOTING........................................10  substrate (board cover or other control tiles) indicates the ma-
                                                                chine has fallen outside an acceptable deviation of .0399.
      FACTORY RECALIBRATION
      EXCHANGE PROGRAM.....................................11
                                                                To achieve accurate results, yourJames Machine must be opera-
      HELP SCREENS...................................................12  ted in a temperature & humidity controlled environment of:
                                                                70º F  (+3º to -5º F) at 50% humidity (+-) 3%.
   1   2   3   4   5   6   7